The Half Bridge instruction should be selected (not the 3 Wire or 4 Wire options) in Short Cut. The following parameters are data logger specific, but the following can be used on the CR1000 datalogger with the sensor wired to SE1, and Vx1:
How many HalfBr Sensors? | 1 |
Total Bridge Resistance | 1000 ohm |
Excitation Voltage | 2500 mV |
Sensors per excitation channel | 1 |
Range of sensor voltage | 2500 mV |
Reverse Excitation | Yes |
Measurement Integration |
50 or 60Hz |
Settling Time | 20000 µs |
Multiplier | Determined during installation (see manual) |
Offset | Determined during installation (see manual) |